MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2008-11-14 for MEDICAL GAS SUPPLY OUTLET manufactured by Allied Health Products, Inc..
[19177402]
Nitric oxide hose was plugged into carbon dioxide outlet. Pt was ventilated with co2 instead of nitric oxide. Anesthesiologist noted that pt's co2 level was 103. At that time, the md noticed the mis-connection.
Patient Sequence No: 1, Text Type: D, B5
Report Number | MW5009016 |
MDR Report Key | 1234564 |
Date Received | 2008-11-14 |
Date of Report | 2008-11-14 |
Date of Event | 2008-10-09 |
Date Added to Maude | 2009-01-06 |
Event Key | 0 |
Report Source Code | Voluntary report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | RISK MANAGER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Generic Name | MEDICAL GAS SUPPLY OUTLET |
Product Code | ECX |
Date Received | 2008-11-14 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ALLIED HEALTH PRODUCTS, INC. |
Manufacturer Address | CHEMOTRON BEACON MEDAES |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Deathisabilit | 2008-11-14 |