COMPAX 40E 45435245 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2008-11-12 for COMPAX 40E 45435245 NA manufactured by Ge Medical Systems, Llc.

Event Text Entries

[994649] It was reported that the table locks did not actuate, causing the tabletop to unexpectedly move in the longitudinal and lateral directions without resistance (free float). There was no injury reported. This situation could contribute to an injury if a patient or operator were unaware of this condition while loading or unloading a patient. The ensuing instability could lead to a fall.
Patient Sequence No: 1, Text Type: D, B5


[8134838] The ge field engineer (fe) was dispatched to the site and found that screws fell out of the bucky tray and shorted the table's circuit breaker, interrupting the power supplied to the table locks. The site contact reported to the fe that the table floated in the longitudinal axis only. The fe visited the site again to confirm the longitudinal float, but found that the table would float in both longitudinal and lateral axes when there was a loss of power to the table locks.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2126677-2008-00091
MDR Report Key1235629
Report Source07
Date Received2008-11-12
Date of Report2008-10-06
Date of Event2008-10-06
Date Mfgr Received2008-11-06
Device Manufacturer Date1994-09-01
Date Added to Maude2009-06-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARTHA KAMROW
Manufacturer Street3000 N GRANDVIEW BLVD. W709
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal53188
Manufacturer Phone2623127196
Manufacturer G1GE MEDICAL SYSTEMS, LLC
Manufacturer Street3000 N GRANDVIEW BLVD.
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal Code53188
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOMPAX 40E
Product CodeIZZ
Date Received2008-11-12
Model Number45435245
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGE MEDICAL SYSTEMS, LLC
Manufacturer Address3000 N GRANDVIEW BLVD. WAUKESHA WI 53188 US 53188


Patients

Patient NumberTreatmentOutcomeDate
10 2008-11-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.