REALTIME HIV-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2008-12-04 for REALTIME HIV-1 manufactured by Abbott Molecular, Inc..

Event Text Entries

[18217177] To conclude, this thermal non-uniformity can result in possible-mis-quantitation too low for low-level samples when using the realtime assay when a fill state of 49 to 88 specimens on the 96-well pcr plate on the m2000rt instrument. Importantly, the low-level samples are accurately quantitated when running a fill state of 48 or 96 wells on the 96-well pcr plate using the realtime hiv-1 assay on the m2000rt instrument.
Patient Sequence No: 1, Text Type: N, H10


[18370585] A determination was made that thermal non-uniformity existed across the 96-well pcr plate when using the m2000rt instrument. This thermal non-uniformity can result in possible mis-quantitation too low for low level samples when running the realtime assay when using a fill state of 49 to 88 specimens.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005248192-2008-00005
MDR Report Key1257470
Report Source07
Date Received2008-12-04
Date of Report2008-12-02
Date of Event2008-11-04
Date Mfgr Received2008-07-23
Date Added to Maude2009-08-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPAULA MARTIN
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2243617333
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameREALTIME HIV-1
Generic NameRT-PCR ASSAY FOR HIV-1 QUANTITATION
Product CodeMTL
Date Received2008-12-04
Returned To Mfg2008-07-23
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer Address1300 EAST TOUHY AVENUE DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2008-12-04