AXP AUTOXPRESS PLATFORM 8-5101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2008-12-08 for AXP AUTOXPRESS PLATFORM 8-5101 manufactured by Thermogenesis Corp..

Event Text Entries

[1020438] A customer call came in on (b)(6)2008, to report that white debris was found in the disposable freezing bag.
Patient Sequence No: 1, Text Type: D, B5


[8130898] We have completed our eval of the returned disposable bagset. The white debris was sent out for fourier transform infrared (ftir) spectroscopy for identification. The particle was identified as bagset stopcock material. This is an isolated event for this lot of disposable bagsets from this mfr. We will continue to monitor and trend. No adverse events have been reported for this occurrence.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2950374-2008-00003
MDR Report Key1258146
Report Source04
Date Received2008-12-08
Date of Report2008-12-08
Date of Event2008-11-11
Date Mfgr Received2008-11-11
Device Manufacturer Date2008-02-01
Date Added to Maude2009-12-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTRISTA MADSEN, MGR
Manufacturer Street2711 CITRUS RD.
Manufacturer CityRANCHO CORDOVA CA 95742
Manufacturer CountryUS
Manufacturer Postal95742
Manufacturer Phone9168585109
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAXP AUTOXPRESS PLATFORM
Product CodeOAO
Date Received2008-12-08
Returned To Mfg2008-11-13
Model Number8-5101
Catalog Number8-5101
Lot Number27066-0000-02
ID NumberNA
Device Expiration Date2009-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTHERMOGENESIS CORP.
Manufacturer Address2711 CITRUS RD. RANCHO CORDOVA CA 95742 US 95742


Patients

Patient NumberTreatmentOutcomeDate
10 2008-12-08

© 2026 FDA.report