HAUSTED VIDEO IMAGING CHAIR 2661765

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2008-11-18 for HAUSTED VIDEO IMAGING CHAIR 2661765 manufactured by Steris Corporation - Montgomery.

Event Text Entries

[990745] While the hospital staff was preparing to transfer a patient from a video imaging chair, the back section of the chair broke and the patient fell.
Patient Sequence No: 1, Text Type: D, B5


[8174098] A steris service technician inspected the chair, which was manufactured in 1995. Inspection of the chair showed extensive corrosion and that the gas spring connected to the chair back section sheared off at the weld where it attached to the seat assembly. The chair was maintained by hospital's maintenance personnel, who indicated that such equipment is not included on a preventive maintenance plan and that they perform maintenance on an "as needed" basis when a unit breaks. The hospital reported that the patient was not injured from the fall.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1043572-2008-00020
MDR Report Key1271751
Report Source05,06,07
Date Received2008-11-18
Date of Report2008-11-17
Date of Event2008-10-15
Date Facility Aware2008-10-15
Report Date2008-11-17
Date Mfgr Received2008-10-16
Device Manufacturer Date1995-01-02
Date Added to Maude2009-11-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROBERT MCCALL
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY AL 36109
Manufacturer CountryUS
Manufacturer Postal36109
Manufacturer Phone3342133150
Manufacturer G1STERIS CORPORATION - MONTGOMERY
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY AL 36109
Manufacturer CountryUS
Manufacturer Postal Code36109
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHAUSTED VIDEO IMAGING CHAIR
Generic NameMANUAL OPERATING CHAIRS AND ACCESSORIES
Product CodeFZK
Date Received2008-11-18
Model Number2661765
Catalog Number2661765
Lot NumberNA
ID NumberVIC429ST
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age156 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - MONTGOMERY
Manufacturer AddressMONTGOMERY AL 36109 US 36109


Patients

Patient NumberTreatmentOutcomeDate
10 2008-11-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.