BREMER 3D HALO CROWN SYSTEM 292301000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,07 report with the FDA on 2008-12-23 for BREMER 3D HALO CROWN SYSTEM 292301000 manufactured by Depuy Spine, Inc..

Event Text Entries

[981489] Initial application of a skull brace was done for dens fracture in 2008. The user noted that when he tightened the pins to set the product on the patient, one of the left posterior pins was loose and easily moved forward. The contact in other country reported that the skull pin had been re-tightened after the initial fitting. The following month, it was found the skull pin had penetrated through the bone. Contact reported no significant patient complication at this time.
Patient Sequence No: 1, Text Type: D, B5


[8168571] Note: this event occurred with a sterile version of the 3d crown that is sold only in other country (catalog number 292301001). As the equivalent device, non-sterile, is sold in the usa as we have filed this mdr using the fda cleared catalog number. No connection could be made between the reported event and any shortcoming of the device. Care must be taken during application and subsequent tightening to ensure that skull pins do not advance through the bone.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1526439-2008-00212
MDR Report Key1275052
Report Source01,05,07
Date Received2008-12-23
Date Mfgr Received2008-11-28
Device Manufacturer Date2008-03-01
Date Added to Maude2009-01-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMATTHEW KING
Manufacturer Street325 PARAMOUNT DR.
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088283106
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameBREMER 3D HALO CROWN SYSTEM
Generic NameHAX/TONG, SKULL FOR TRACTION
Product CodeHAX
Date Received2008-12-23
Model NumberNA
Catalog Number292301000
Lot Number060395
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SPINE, INC.
Manufacturer Address325 PARAMOUNT DR. RAYNHAM MA 02767 US 02767


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2008-12-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.