BUCKY/ACL-110 VOLT SIZE WALL BUCKY 249112SL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2009-01-07 for BUCKY/ACL-110 VOLT SIZE WALL BUCKY 249112SL manufactured by Liebel - Flarsheim Co..

Event Text Entries

[980865] Customer reports that during automatic ejection, the 10 x 12 cassette does not stay in place, but "falls to the floor. "
Patient Sequence No: 1, Text Type: D, B5


[8171300] Liebel flarsheim manufacturing report: liebel flarsheim was not able to reproduce the problem with the two types of 10" x 12" cassettes that we have in-house. In discussions with the technical services and the product engineer, we could only surmise that the cassete's forward motion was continuing after the cassette release, propelling it to the floor. However, the 'rails' in which the cassette rides would still be engaged and not allow the cassette to tip-out even if the cassette was half way out of the opening. The only way that the rails would not hold the cassette would be if the entry sensor fails, causing the rails to return to their home position. It was also suggested that by taping the cassette edges, you can increase rail friction, or to suggest discontinue the use of the automatic eject mode as means to eliminate the problem.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1518293-2008-00044
MDR Report Key1278961
Report Source06
Date Received2009-01-07
Date of Report2003-04-07
Date Mfgr Received2003-04-07
Device Manufacturer Date2001-05-01
Date Added to Maude2009-04-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDAVID BENSON
Manufacturer Street2111 EAST GALBRAITH RD
Manufacturer CityCINCINNATI OH 45237
Manufacturer CountryUS
Manufacturer Postal45237
Manufacturer Phone5139485719
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBUCKY/ACL-110 VOLT SIZE
Generic NameCASSETTE WALL BUCKY SYSTEM
Product CodeIXA
Date Received2009-01-07
Model NumberWALL BUCKY
Catalog Number249112SL
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIEBEL - FLARSHEIM CO.