TX-1 POWER TRACTION DEVICE 7240

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2009-01-02 for TX-1 POWER TRACTION DEVICE 7240 manufactured by Chattanooga Group.

Event Text Entries

[17017783] A pt was injured during traction therapy and treatment was stopped. After the event, a tech was sent and determined that the unit was out of calibration by 20-22%. He recalibrated the unit. The pt returned to the p. T. Office in 2006 inflamed and swollen. She is scheduled for physical therapy about four days later.
Patient Sequence No: 1, Text Type: D, B5


[17242327] Only control panel (p/n 70217) was returned for evaluation, which indicates that at the time of treatment, the unit was 1. 5 years old. It is not known at this time whether the unit was calibrated after 1 year of service or had undergone any preventative maintenance. Calibration was checked: 20 lbs= 22 lbs, 100 lbs= 108 lbs, 200 lbs= 214 lbs. Unit was not properly calibrated by field technician. Meter would only go up to 198 lbs max. , zero was flashing -01. Conclusion: control board was removed from unit and recalibrated. It is impossible to determine the state of calibration at the time of removal from unit. The calibration done by the field technician was not correct. Unit calibrated properly and returned to stock.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1022819-2008-00330
MDR Report Key1280492
Report Source06
Date Received2009-01-02
Date of Report2006-09-25
Date of Event2006-09-25
Date Mfgr Received2006-09-25
Date Added to Maude2009-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL TREAS
Manufacturer Street4717 ADAMS ROAD
Manufacturer CityHIXSON TN 37343
Manufacturer CountryUS
Manufacturer Postal37343
Manufacturer Phone4238707207
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTX-1 POWER TRACTION DEVICE
Product CodeITH
Date Received2009-01-02
Returned To Mfg2006-10-26
Model Number7240
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHATTANOOGA GROUP
Manufacturer Address4717 ADAMS ROAD HIXSON TN 37343 US 37343


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2009-01-02