PALL BACTERIA DETECTION SYSTEM FOR PLATELET TRANSFUSION PRODUCTS BDS02

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2009-01-09 for PALL BACTERIA DETECTION SYSTEM FOR PLATELET TRANSFUSION PRODUCTS BDS02 manufactured by Ensatec, S.a. De C.v..

Event Text Entries

[21399377] It was reported from a blood center that one (1) apparently false negative result was obtained while using the pall bacterial detection device; i. E. An apheresis platelet unit sample was tested for % oxygen and measured within specifications for a pass result. This apheresis unit was split into two products and sent to two different facilities. First split unit was sent to a transfusion facility that maintains an sop directing the culture of a sample of each platelet product when issued for transfusion. As such, the 1st split unit sample was cultured and tested positive for strep viridans. The 2nd split unit had been sent from the blood center to a different transfusion facility that does not routinely culture samples of platelets prior to transfusion. Both platelet product split units were administered to different patients, one at each transfusion, and no transfusion reactions were reported from either facility.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617787-2009-00006
MDR Report Key1282104
Report Source05,06,07
Date Received2009-01-09
Date of Report2008-12-30
Date of Event2008-12-29
Date Facility Aware2008-12-30
Report Date2008-12-30
Date Reported to Mfgr2008-12-30
Date Mfgr Received2008-12-30
Date Added to Maude2009-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. LEONARD BERMAN
Manufacturer Street25 HARBOR PARK DR
Manufacturer CityPORT WASHINGTON NY 11050
Manufacturer CountryUS
Manufacturer Postal11050
Manufacturer Phone5168019183
Manufacturer G1ENSATEC, S.A. DE C.V.
Manufacturer StreetFRACCIONAMIENTO EL FLORIDO CALLE COLINAS #11730
Manufacturer CityTIJUANA, B.C.,, 22680
Manufacturer CountryMX
Manufacturer Postal Code22680
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePALL BACTERIA DETECTION SYSTEM FOR PLATELET TRANSFUSION PRODUCTS
Generic NameBACTERIA DETECTIN SYSTEM
Product CodeMZC
Date Received2009-01-09
Model NumberBDS02
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerENSATEC, S.A. DE C.V.
Manufacturer AddressFRACCIONAMIENTO EL FLORIDO CALLE COLINAS #11730 TIJUANA, B.C.,, 22680 MX 22680


Patients

Patient NumberTreatmentOutcomeDate
10 2009-01-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.