DIAMOND CONNECTOR, N2O *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2008-11-20 for DIAMOND CONNECTOR, N2O * manufactured by Ohio Medical Corporation.

Event Text Entries

[998593] We discovered a problem for hospitals using wall provided n2o and co2 gas. We are able to plug the n2o ohio diamond connector into the chemtron model 500 n20 outlet, which it is intended for, but also into the co2 outlet, which it is not intended for. The ohio diamond connector for co2 does not fit in any other gas outlet. The outlets are designed to accept either a chemtron or ohio diamond connector. Labeling the connectors will not help, as the problem can still occur if the connector is turned over and inserted into the wall outlet. We plan to change the n2o connector from ohio diamond to chemtron on all anesthesia machines. The chemtron fitting for n2o can not fit into the co2 wall outlet. ====================== manufacturer response for n2o and co2 connectors, (brand not provided)======================hospital contacted both ohio medical and allied health care to alert them of problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1282626
MDR Report Key1282626
Date Received2008-11-20
Date of Report2008-11-20
Date of Event2008-11-13
Report Date2008-11-20
Date Reported to FDA2008-11-20
Date Added to Maude2009-01-12
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameDIAMOND CONNECTOR, N2O
Generic NameCONNECTOR, TUBING
Product CodeCAM
Date Received2008-11-20
Model Number*
Catalog Number*
Lot Number*
ID Number*
OperatorOTHER
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerOHIO MEDICAL CORPORATION
Manufacturer AddressDIRECTOR OF REGULATORY AFFAIRS 1111 LAKESIDE DRIVE GURNEY IL 60031 US 60031

Device Sequence Number: 2

Brand NameCHEMETRON WALL OUTLET
Generic NameWALL OUTLET, GAS
Product CodeCAM
Date Received2008-11-20
Model Number500 SERIES
Catalog Number*
Lot Number*
ID Number*
Device Sequence No2
Device Event Key0
ManufacturerALLIED HEALTHCARE PRODUCTS, INC.
Manufacturer Address1720 SUBLETTE AVENUE ST. LOUIS OH * US *


Patients

Patient NumberTreatmentOutcomeDate
10 2008-11-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.