COONRAD/MORREY TOTAL ELBOW PIN/BUSHING REPLACEMENT KIT SMALL 32810502501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,08 report with the FDA on 2008-12-08 for COONRAD/MORREY TOTAL ELBOW PIN/BUSHING REPLACEMENT KIT SMALL 32810502501 manufactured by Zimmer, Inc..

Event Text Entries

[1039533] It is reported that when the surgeon tried to implant the replacement kit, it would not fit. It was reported the elbow was a size small. Another replacement kit was opened, which was implanted successfully.
Patient Sequence No: 1, Text Type: D, B5


[8168629] Evaluation summary: the complaint does not state whether or not the second kit used was the same item number. No item numbers of the previously implanted stems were provided. X-rays were not returned for evaluation. The complaint indicates that the elbow was small. There are different revision bushing kits for the various combinations of implant stems, as noted in the surgical technique. Based on the information provided, the exact cause can not be determined with certainty. Device history records indicate all components were manufactured and inspected to specification.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1822565-2008-00885
MDR Report Key1310239
Report Source05,08
Date Received2008-12-08
Date of Report2008-11-03
Date of Event2008-10-31
Date Facility Aware2008-11-03
Report Date2008-11-03
Date Mfgr Received2008-11-06
Device Manufacturer Date2008-06-01
Date Added to Maude2009-05-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBRIAN FIEDLER
Manufacturer StreetP.O. BOX 708
Manufacturer CityWARSAW IN 465810708
Manufacturer CountryUS
Manufacturer Postal465810708
Manufacturer Phone8006136131
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOONRAD/MORREY TOTAL ELBOW PIN/BUSHING REPLACEMENT KIT SMALL
Generic NameEXTREMITY PROSTHESIS
Product CodeKWJ
Date Received2008-12-08
Returned To Mfg2008-11-03
Model NumberNA
Catalog Number32810502501
Lot Number60996790
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER, INC.
Manufacturer AddressP.O. BOX 708 WARSAW IN 46581070 US 46581 0708


Patients

Patient NumberTreatmentOutcomeDate
10 2008-12-08

© 2026 FDA.report