THERATHERM MOIST HEAT 1032

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-02-24 for THERATHERM MOIST HEAT 1032 manufactured by Shian Ja Meei Enterprise Co., Ltd..

Event Text Entries

[1006341] Patient was using the theratherm electric moist heating pad. She claimed the product timer turned it off, and she laid it on her leather sofa. When she came back into the room, she claimed the heating pad had turned itself on, and was smoldering the sofa.
Patient Sequence No: 1, Text Type: D, B5


[8220950] The controller functions were tested with no deficiencies noted. Unit shut off when set time expires and the display goes blank, pad does not continue to heat and does not turn back on. Controller case was examined, and was not damaged. I. R. Temperature of controller was evaluated against controlled temperatures and no deficiencies were noted. The controller shuts power to heater off at 127 f. Controller was cycled several times and no malfunctions occur. Both sides of the outer cover are scorched and blackened near the sewn on label cord from controller to the pad is discolored a dirty pinkish hue. While no evidence of malfunction was identified, unit does appear to have undergone a scorch situation so mdr filed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1022819-2008-00381
MDR Report Key1322456
Report Source05
Date Received2009-02-24
Date of Report2007-04-23
Date Mfgr Received2007-04-23
Date Added to Maude2009-02-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL TREAS
Manufacturer Street4717 ADAMS ROAD
Manufacturer CityHIXSON TN 37343
Manufacturer CountryUS
Manufacturer Postal37343
Manufacturer Phone4238707207
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHERATHERM MOIST HEAT
Generic NameIMA
Product CodeIMA
Date Received2009-02-24
Returned To Mfg2007-04-30
Model Number1032
OperatorPATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSHIAN JA MEEI ENTERPRISE CO., LTD.
Manufacturer AddressTW


Patients

Patient NumberTreatmentOutcomeDate
10 2009-02-24

© 2026 FDA.report