RX ERCP CANNULA TAPERED TIP M00545770 4577

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2009-02-25 for RX ERCP CANNULA TAPERED TIP M00545770 4577 manufactured by Boston Scientific Corporation-spencer.

Event Text Entries

[1004920] It was reported to boston scientific corporation that a rx ercp cannula tapered tip device was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure (patient age, gender and weight unknown). According to the complainant, during the procedure, upon flushing, the contrast started to leak out of the catheter 25 centimeters from the distal tip. Another rx ercp cannula tapered tip device was used to complete the procedure. There were no patient complications reported and the patient's condition at the conclusion of the procedure was reported to be "fine. "
Patient Sequence No: 1, Text Type: D, B5


[8153566] A visual evaluation found no visual defects with the device. The outer diameter of the joint was within the manufacturing specification. A functional evaluation was performed by injecting water through the device using a 20-milliliter syringe. The device did not leak while injecting water through it. A lot history search was performed and found no other complaints against this lot.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005099803-2008-02719
MDR Report Key1324909
Report Source05,07
Date Received2009-02-25
Date of Report2007-06-01
Date of Event2007-06-01
Date Mfgr Received2007-06-01
Device Manufacturer Date2007-04-23
Date Added to Maude2009-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKEN HIRAKAWA
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086836264
Manufacturer G1BOSTON SCIENTIFIC CORPORATION-SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRX ERCP CANNULA TAPERED TIP
Generic NameENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) CANNULA
Product CodeODD
Date Received2009-02-25
Returned To Mfg2008-08-30
Model NumberM00545770
Catalog Number4577
Lot Number0009576379
Device Expiration Date2009-04-17
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION-SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2009-02-25