RX ERCP CANNULA TAPERED TIP M00545770 4577

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,05,07 report with the FDA on 2009-03-05 for RX ERCP CANNULA TAPERED TIP M00545770 4577 manufactured by Boston Scientific Corporation-spencer.

Event Text Entries

[1066636] It was reported to boston scientific corporation that a rx ercp cannula tapered tip was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure (patient age, gender and weight are unknown). According to the complainant, this device was inserted along with the guide wire. It was removed and then reinserted without any difficulty. When attempting to remove it again, it appeared to be stuck and it had to be removed as one unit. After removal, a pinhole tear was observed on the cannula. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be "good. "
Patient Sequence No: 1, Text Type: D, B5


[8221113] A visual examination of the returned device noted there is damage in the joint section of the sheath. A functional examination was performed using a. 035 guide wire. The guide wire was advanced from the guide wire port to the damaged joint section of the sheath and could not be advanced any further. The complaint has been confirmed; the root cause of the broken joint is undetermined. A review of the device history record was performed and no issues related to this complaint were noted. A review of complaint history for this lot revealed there are no other complaints against this lot number.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005099803-2008-03087
MDR Report Key1333838
Report Source00,01,05,07
Date Received2009-03-05
Date of Report2006-01-10
Date of Event2006-01-09
Date Mfgr Received2006-01-10
Device Manufacturer Date2005-11-16
Date Added to Maude2009-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKEN HIRAKAWA
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086836264
Manufacturer G1BOSTON SCIENTIFIC CORPORATION-SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRX ERCP CANNULA TAPERED TIP
Generic NameENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) CANNULA
Product CodeODD
Date Received2009-03-05
Returned To Mfg2006-02-22
Model NumberM00545770
Catalog Number4577
Lot Number0008163967
Device Expiration Date2007-11-05
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION-SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2009-03-05