MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2009-03-12 for AXSYM RUBELLA LGG ANTIBODY 3B23-20 manufactured by Abbott Laboratories.
[1079548]
The customer contaced abbott regarding three attempts to calibrate the axsym rubella lgg assay which failed, and error code 1111 (calibration check failure, m-cal 1, response too large) was generated. The customer reported axsym maintenance was up to date. For troubeshooting, the customer attempted calibration using water instead of master calibrator 1, which was successful. The customer cleaned the instrument optics and attempted recalibration which failed again with the same error message. The customer was sent new reagents and calibrators and upon recalibration with the new material, the same callibration failure persisted. The customer was advised to centrifuge calibrator a, then use the centrifuged calibrator in the rubella recalibration which was successful. There was no impact to patient management reported by the customer.
Patient Sequence No: 1, Text Type: D, B5
[8253867]
This is an initial report. An investigation is in process. A final report will be submitted when the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1415939-2008-00288 |
MDR Report Key | 1341434 |
Report Source | 01 |
Date Received | 2009-03-12 |
Date of Report | 2008-01-31 |
Date of Event | 2008-01-30 |
Date Mfgr Received | 2008-06-02 |
Date Added to Maude | 2009-03-12 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | KHAUDEJA BANO |
Manufacturer Street | 100 ABBOTT PARK ROAD |
Manufacturer City | ABBOTT PARK IL 600646112 |
Manufacturer Country | US |
Manufacturer Postal | 600646112 |
Manufacturer Phone | 8479386319 |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Removal Correction Number | 1415939-6/06/08-004-C |
Event Type | 3 |
Type of Report | 3 |
Brand Name | AXSYM RUBELLA LGG ANTIBODY |
Generic Name | MEIA FOR THE MEASUREMENT OF LGG ANTIBODY TO THE RUBELLA VIRUS |
Product Code | GOL |
Date Received | 2009-03-12 |
Model Number | NA |
Catalog Number | 3B23-20 |
Lot Number | 56323M101 |
ID Number | UNK |
Device Expiration Date | 2008-06-30 |
Operator | OTHER |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT LABORATORIES |
Manufacturer Address | 100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2009-03-12 |