MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2009-03-12 for AXSYM RUBELLA LGG ANTIBODY 3B23-20 manufactured by Abbott Laboratories.
[1072326]
The customer reported four failed calibrations of the abbott axsym rubella lgg assay, where error message 1018 (calibration check failure, cal a, result too high) was generated. Abbott requested the customer print out the failed and the last successful rubella calibration and fax for evaluation by abbott. The customer's data revealed elevated values for the calibrators. The customer reported the axsym probes and meia lamp were recently checked. The customer was sent a new meia lamp and new lot of rubella reagent and calibrators. Recalibration was successful with the new rubella reagent. There was no impact to patient management reported by the customer.
Patient Sequence No: 1, Text Type: D, B5
[8222926]
(no consequence or impact to patient), (calibration error), (error message given) this is an initial report. An investigation is in process. A final report will be submitted when the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1415939-2008-00284 |
MDR Report Key | 1341450 |
Report Source | 01 |
Date Received | 2009-03-12 |
Date of Report | 2008-01-17 |
Date of Event | 2008-01-17 |
Date Mfgr Received | 2008-06-02 |
Date Added to Maude | 2009-03-12 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | KHAUDEJA BANO |
Manufacturer Street | 100 ABBOTT PARK ROAD |
Manufacturer City | ABBOTT PARK IL 600646112 |
Manufacturer Country | US |
Manufacturer Postal | 600646112 |
Manufacturer Phone | 8479386319 |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Removal Correction Number | 1415939-6/06/08-004-C |
Event Type | 3 |
Type of Report | 3 |
Brand Name | AXSYM RUBELLA LGG ANTIBODY |
Generic Name | MEIA FOR THE MEASUREMENT OF LGG ANTIBODY TO THE RUBELLA VIRU |
Product Code | GOL |
Date Received | 2009-03-12 |
Model Number | NA |
Catalog Number | 3B23-20 |
Lot Number | 55922M101 |
ID Number | NA |
Device Expiration Date | 2008-06-20 |
Operator | OTHER |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT LABORATORIES |
Manufacturer Address | 100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2009-03-12 |