MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2009-03-12 for AXSYM RUBELLA LGG ANTIBODY 3B23-20 manufactured by Abbott Laboratories.
        [1072327]
The customer stated that the axsym rubella lgg reagent calibration failed twice with an error code on the axsym analyzer that indicated the m-cal 1 response was too large. The customer received a new calibrator lot and recalibrated successfully. No impact to patient management reported.
 Patient Sequence No: 1, Text Type: D, B5
        [8223444]
(no consequence or impact to patient). (calibration error). (error message given) this is an initial report. An investigation is in process. A final report will be submitted when the investigation is complete.
 Patient Sequence No: 1, Text Type: N, H10
| Report Number | 1415939-2008-00266 | 
| MDR Report Key | 1341452 | 
| Report Source | 01,05 | 
| Date Received | 2009-03-12 | 
| Date of Report | 2008-04-21 | 
| Date of Event | 2008-04-18 | 
| Date Mfgr Received | 2008-06-02 | 
| Date Added to Maude | 2009-03-12 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 0 | 
| Event Location | 0 | 
| Manufacturer Contact | KHAUDEJA BANO | 
| Manufacturer Street | 100 ABBOTT PARK ROAD | 
| Manufacturer City | ABBOTT PARK IL 600646112 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 600646112 | 
| Manufacturer Phone | 8479386319 | 
| Single Use | 3 | 
| Remedial Action | OT | 
| Previous Use Code | 3 | 
| Removal Correction Number | 1415939-6/06/08-004-C | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | AXSYM RUBELLA LGG ANTIBODY | 
| Generic Name | MEIA FOR THE MEASUREMENT OF LGG ANTIBODY TO THE RUBELLA VIRUS | 
| Product Code | GOL | 
| Date Received | 2009-03-12 | 
| Model Number | NA | 
| Catalog Number | 3B23-20 | 
| Lot Number | 56659M201 | 
| ID Number | NA | 
| Device Expiration Date | 2008-07-21 | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | Y | 
| Device Eval'ed by Mfgr | Y | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | ABBOTT LABORATORIES | 
| Manufacturer Address | 100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2009-03-12 |