MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2009-03-12 for AXSYM RUBELLA LGG ANTIBODY 3B23-20 manufactured by Abbott Laboratories.
[20942092]
The customer reported three failed axsym rubella lgg calibrators, where error code 1048 (calibration check failure, calibrators a/b ratio too high) was generated. Another failed attempted calibration was performed with a different calibrator lot where error code 1111 (calibration check failure, master calibrator 1, result too high) was generated. The customer was sent another lot of reagent and abbott requested faxes of the calibration data for evaluation. The customer reported the axsym probes were recently replaced. Upon follow up with the customer after new lot of rubella reagent was sent, the customer reported everything had returned to normal. Additionally the customer reported there was no impact to patient management.
Patient Sequence No: 1, Text Type: D, B5
[21201731]
This is an initial report. An investigation is in process. A final report will be submitted when the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1415939-2008-00277 |
MDR Report Key | 1341463 |
Report Source | 01 |
Date Received | 2009-03-12 |
Date of Report | 2007-12-06 |
Date of Event | 2007-12-05 |
Date Mfgr Received | 2008-06-02 |
Date Added to Maude | 2009-03-12 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | KHAUDEJA BANO |
Manufacturer Street | 100 ABBOTT PARK ROAD |
Manufacturer City | ABBOTT PARK IL 600646112 |
Manufacturer Country | US |
Manufacturer Postal | 600646112 |
Manufacturer Phone | 8479386319 |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Removal Correction Number | 1415939-6/06/08-004-C |
Event Type | 3 |
Type of Report | 3 |
Brand Name | AXSYM RUBELLA LGG ANTIBODY |
Generic Name | MEIA FOR THE MEASUREMENT OF LGG ANTIBODY TO THE RUBELLA VIRUS |
Product Code | GOL |
Date Received | 2009-03-12 |
Model Number | NA |
Catalog Number | 3B23-20 |
Lot Number | 53587M101 |
ID Number | NA |
Device Expiration Date | 2008-04-15 |
Operator | OTHER |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT LABORATORIES |
Manufacturer Address | 100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2009-03-12 |