CO2 CONNET 2 CHEMETRON OUTLET 64-02-5007

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2009-01-23 for CO2 CONNET 2 CHEMETRON OUTLET 64-02-5007 manufactured by Allied Healthcare Products, Inc..

Event Text Entries

[1070450] It was reported that the hospital using wall provided n2o and co2 gas had a problem. The hospital was able to plug the n2o ohio diamond connector into the chemetron model 500 connect 2 n2o outlet, which it is intended for, but also into the co2 outlet, which it is not intended for. The ohio diamond connector for co2 does not fit in any other gas outlet. The outlets are designed to accept either a chemetron or ohio diamond connector. Labeling the connectors will not help, as the problem can still occur if the connector is turned over and inserted into the wall outlet. We plan to change the n2o connector from ohio diamond to chemetron on all anesthesia machines. The chemetron fitting for n2o can not fit into the co2 wall outlet.
Patient Sequence No: 1, Text Type: D, B5


[8174897] Connect 2 outlets are labeled for the gas they are intended for. (carbon dioxide). Connect 2 outlets are color coded for the gas they are intended for. (gray for carbon dioxide). The only what to make an ohio n2o adapter go into a chemetron co2 connect 2 outlet is to turn it upside down (180 degrees).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1924066-2009-00001
MDR Report Key1342634
Report Source05,06
Date Received2009-01-23
Date of Report2009-01-23
Date of Event2008-11-13
Date Mfgr Received2009-01-20
Date Added to Maude2010-01-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional0
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer Street1720 SUBLETTE AVE.
Manufacturer CityST. LOUIS MO 63110
Manufacturer CountryUS
Manufacturer Postal63110
Manufacturer Phone3147712400
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCO2 CONNET 2 CHEMETRON OUTLET
Generic NameCO2 OUTLET STATION
Product CodeCAM
Date Received2009-01-23
Model Number64-02-5007
Catalog Number64-02-5007
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALLIED HEALTHCARE PRODUCTS, INC.
Manufacturer AddressST. LOUIS MO 63110 US 63110


Patients

Patient NumberTreatmentOutcomeDate
10 2009-01-23

© 2026 FDA.report