LOW MAIN TURBINE FOR STAR 430 PB HANDPIECE 102-4535

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2006-09-14 for LOW MAIN TURBINE FOR STAR 430 PB HANDPIECE 102-4535 manufactured by Dental Instruments.

Event Text Entries

[1079669] The doctor was performing crown and bridge work on a patient, when the bur walked out of the chuck, and was swallowed by the patient. The patient required two colonoscopy procedures to surgically remove the swallowed bur.
Patient Sequence No: 1, Text Type: D, B5


[8190230] The doctor advised us that he had noticed the bur continued to slip prior to this event, and continued to use it without notifying us of the potential problem.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2437811-2006-00001
MDR Report Key1348644
Report Source05
Date Received2006-09-14
Date of Report2006-08-15
Date of Event2006-05-18
Date Mfgr Received2006-08-15
Date Added to Maude2009-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street137 COMMERCIAL ST., SUITE 302
Manufacturer CityPLAINVIEW NY 11803
Manufacturer CountryUS
Manufacturer Postal11803
Manufacturer Phone5167192101
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLOW MAIN TURBINE FOR STAR 430 PB HANDPIECE
Generic NameDENTAL
Product CodeEJB
Date Received2006-09-14
Returned To Mfg2006-09-05
Catalog Number102-4535
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDENTAL INSTRUMENTS
Manufacturer AddressPLAINVIEW NY US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2006-09-14

© 2026 FDA.report