COOL FLOW IRRIGATION PUMP M-5491-02 CFP002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2009-04-01 for COOL FLOW IRRIGATION PUMP M-5491-02 CFP002 manufactured by Hei, Inc..

Event Text Entries

[1069661] It was reported that during the procedure, the customer experienced high impedance readings on the stockert generator while using the hansen system. The maximum valves during the lesion were 30 watts, 147 ohms, and 48 degrees celsius. The rf energy was terminated when the impedance reading increased to 150 ohms. While customer was troubleshooting the high impedance readings, the patient's blood pressure dropped precipitously. The procedure was stopped and a pericardiocentesis drain was placed. The patient was ultimately transferred to the or where surgical repair of the basal left atrium was performed. Additional information obtained: the patient did well and fully recovered after surgery.
Patient Sequence No: 1, Text Type: D, B5


[8223047] The biosense webster field service engineer (fse) contacted the customer with regards to this incident. The customer was of the opinion that the cool flow pump did not cause the adverse event, and it did not need to be evaluated by biosense webster. Customer used the cool flow pump in a subsequent procedure without any issues.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1721752-2009-00005
MDR Report Key1356312
Report Source06
Date Received2009-04-01
Date of Report2009-03-02
Date of Event2009-03-02
Date Mfgr Received2009-03-02
Date Added to Maude2009-04-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactGAIL HAPNER
Manufacturer Street15715 ARROW HIGHWAY
Manufacturer CityIRWINDALE CA 91706
Manufacturer CountryUS
Manufacturer Postal91706
Manufacturer Phone9098398938
Manufacturer G1HEI, INC.
Manufacturer Street4801 NORTH 63RD ST.
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal Code80301
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOOL FLOW IRRIGATION PUMP
Generic NameLPB PUMP, CIRCULATING-FLUID, LOCALIZED HEAT
Product CodeKQT
Date Received2009-04-01
Model NumberM-5491-02
Catalog NumberCFP002
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerHEI, INC.
Manufacturer AddressBOULDER CO US


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening; 2. Required No Informationntervention 2009-04-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.