S/5 ANESTHESIA MONITOR E-CIAO-00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2009-04-28 for S/5 ANESTHESIA MONITOR E-CIAO-00 manufactured by Ge Healthcare Finland Oy.

Event Text Entries

[1205948] The customer reported that during a case, the physicians on call recognized that they had no gas analysis on the s/5 monitor. This was due to the sample line of the gas analysis module being inadvertently attached to the calibration port. As a result, the physicians were unable to monitor end tidal co2 or oxygen, and were unable to measure the potency of gas delivery. The first issue was well managed as only oxygen was being delivered and the respiratory rate was known. The second issue resulted in the pt receiving twice the amount of anesthetic as intended. While the correct values were displayed on the monitor in the gas settings area, the caregivers did not recognize that sevoflurane was being delivered at 2% because the area that providers are used to referencing (end tidal concentrations) was not functioning. The physicians proceeded with manual ventilation and oxygen. Iv anesthetics were used to keep pt under anesthesia. Ge healthcare's investigation into the reported occurrence is still ongoing. A follow-up report will be issued when the investigation has been completed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610105-2009-00021
MDR Report Key1371272
Report Source99
Date Received2009-04-28
Date of Report2009-04-28
Date of Event2009-04-09
Date Mfgr Received2009-04-09
Device Manufacturer Date2008-08-01
Date Added to Maude2011-03-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANN LEBAR
Manufacturer Street3000 N. GRANDVIEW BLVD MAIL CODE - W450
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal53188
Manufacturer Phone2625482757
Manufacturer G1GE HEALTHCARE FINLAND OY
Manufacturer CityHELSINKI
Manufacturer CountryFI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameS/5 ANESTHESIA MONITOR
Generic NamePARAMETER MODULES
Product CodeCCL
Date Received2009-04-28
Model NumberE-CIAO-00
ID NumberE-CAIO
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerGE HEALTHCARE FINLAND OY
Manufacturer AddressHELSINKI FI


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2009-04-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.