COBAS INTEGRA 400 PLUS 3245233001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2009-02-10 for COBAS INTEGRA 400 PLUS 3245233001 manufactured by Roche Diagnostics.

Event Text Entries

[17369255] Several negative samples of urine tested for benzodiazepine which contained considerable quantities of lorazepam when determined by chromatography. One example provided found 890 ng/ml by chromatography and only 10 ng/ml on this analyzer. Total number of pts involved was not provided. No information was provided to determine if erroneous results were reported or if pts were adversely affected.
Patient Sequence No: 1, Text Type: D, B5


[17667869] Lorazepam appears in the urine largely as the glucoronidated metabolite. It was noted the lorazepam in the specimen was glucoronidated. Cross reactivity to glucoronidated lorazepam is unknown. If additional information is received, appropriate notification will be provided.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2009-01040
MDR Report Key1371284
Report Source05,06
Date Received2009-02-10
Date of Report2009-02-10
Date of Event2009-01-20
Date Facility Aware2009-01-20
Report Date2009-01-20
Date Mfgr Received2009-01-20
Date Added to Maude2009-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactERIC KOLODZIEJ
Manufacturer Street9115 HAGUE RD
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175212834
Manufacturer G1ROCHE INSTRUMENT CENTER AG
Manufacturer StreetTEGIMENTA FORRENSTRASSE
Manufacturer CityROTKREUZ 312-8504
Manufacturer CountrySZ
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOBAS INTEGRA 400 PLUS
Generic NameCLINICAL CHEMISTRY ANALYZER - JJE
Product CodeMGX
Date Received2009-02-10
Catalog Number3245233001
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer AddressINDIANAPOLIS IN US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-02-10

© 2026 FDA.report