QUANTUM THIN HIGH ADD RGP MULTIFOCAL NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-04-29 for QUANTUM THIN HIGH ADD RGP MULTIFOCAL NA manufactured by Unilens Corp., Usa.

Event Text Entries

[14872423] Documentation received with returned product indicated "this lens is defective. Abrasion 2009 rec'd new cl". The following month, the attending healthcare practitioner responded to queries and indicated the abrasion was noted at the time of contact lens dispensing, with the lens being too uncomfortable and painful for the patient. Prophylactic antibiotics were given. No prognosis was given. The pt is not currently wearing a replacement contact lens. No solutions were dispensed to the pt, and the pt is using a deep well screw top contact lens container.
Patient Sequence No: 1, Text Type: D, B5


[15453876] All tested parameters met inspection/release criteria for the returned lens. In the cosmetic examination of the lens surfaces, a small burnished area was found near the edge of the lens that was within acceptable criteria and would not have caused the noted adverse event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1034196-2009-00001
MDR Report Key1373506
Report Source05
Date Received2009-04-29
Date of Report2009-04-29
Date Mfgr Received2009-03-31
Device Manufacturer Date2009-03-01
Date Added to Maude2009-05-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactALAN FRAZER, DIRECTOR
Manufacturer Street10431 72ND ST. NORTH
Manufacturer CityLARGO FL 33777
Manufacturer CountryUS
Manufacturer Postal33777
Manufacturer Phone7275442531
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameQUANTUM THIN HIGH ADD RGP MULTIFOCAL
Generic NameRIGID GAS PERMEABLE CONTACT LENS
Product CodeHPX
Date Received2009-04-29
Returned To Mfg2009-03-31
Model NumberQUANTUM THIN
Catalog NumberNA
Lot Number231093
ID Number7.35 9.2 -8.25
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerUNILENS CORP., USA
Manufacturer AddressLARGO FL 33777 US 33777


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2009-04-29

© 2026 FDA.report