ELITE & USA COLD KNIFE, SERRATED BLADE K-SE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2009-04-16 for ELITE & USA COLD KNIFE, SERRATED BLADE K-SE manufactured by Gyrus Medical Inc..

Event Text Entries

[22050583] The complaint has been confirmed. The electrode has been returned with the knife blade fractured off. The blade fractured at the distal end of the electrode. All parts are accounted for. The weld site is intact. The blade has a slight bend to the proximal end at the fracture point. This would indicate that a sideways force was applied to the blade prior to it fracturing off. The blade of the electrode showed signs of excessive force bending it, the user caused the failure mode.
Patient Sequence No: 1, Text Type: N, H10


[22103124] During a surgical procedure for a right ureteral stricture and ureteral stent, the knife blade on the cold knife broke off and fell into the patient's bladder. All pieces were recovered with no harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183680-2009-00029
MDR Report Key1374599
Report Source06
Date Received2009-04-16
Date of Report2009-04-16
Date of Event2009-03-31
Date Facility Aware2009-03-31
Date Mfgr Received2009-03-31
Date Added to Maude2010-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTERRENCE SULLIVAN
Manufacturer Street136 TURNPIKE RD.
Manufacturer CitySOUTHBOROUGH MO 01772
Manufacturer CountryUS
Manufacturer Postal01772
Manufacturer Phone5088042739
Manufacturer G1GYRUS MEDICAL INC.
Manufacturer Street6655 WEDGWOOD ROAD
Manufacturer CityMAPLE GROVE MN 55311360
Manufacturer CountryUS
Manufacturer Postal Code55311 3602
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELITE & USA COLD KNIFE, SERRATED BLADE
Generic NameCOLD KNIFE, SERRATED BLADE, 3/PKG.
Product CodeEZO
Date Received2009-04-16
Returned To Mfg2009-04-13
Model NumberK-SE
Catalog NumberK-SE
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS MEDICAL INC.
Manufacturer AddressMAPLE GROVE MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-04-16