HANDLES & CONNECTOR WITH DISCHARGE CONTROL FOR INTERNAL DEFIBRILLATION 805249-06

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2009-03-06 for HANDLES & CONNECTOR WITH DISCHARGE CONTROL FOR INTERNAL DEFIBRILLATION 805249-06 manufactured by Physio-control, Inc..

Event Text Entries

[19531358] While picking random samples from different lots for testing, physio-control engineering discovered damaged coating on 10 internal paddles. There have been no field complaints and no service reports during this period. There is an impact to the intended internal paddles device performance. Current flowing from the defect can cause local tissue damage due to thermal heating or due to electroporation. Current flowing from the back of the paddle shunts current from the intended therapeutic path, rendering defibrillation ineffective. This analysis indicates that harm may be caused to the patent due to thermal heating and electroporation. Current shunting is small enough to preclude any significant impact on shock efficacy.
Patient Sequence No: 1, Text Type: D, B5


[19665099] Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803. 56.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3015876-2009-00270
MDR Report Key1377609
Report Source06,07
Date Received2009-03-06
Date of Report2009-02-06
Date of Event2009-02-06
Date Facility Aware2009-02-06
Report Date2009-02-06
Date Mfgr Received2009-02-06
Date Added to Maude2009-07-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactANA SZUCS
Manufacturer Street11811 WILLOWS RD. N.E. PO BOX 97006
Manufacturer CityREDMOND WA 980739706
Manufacturer CountryUS
Manufacturer Postal980739706
Manufacturer Phone4258674000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHANDLES & CONNECTOR WITH DISCHARGE CONTROL FOR INTERNAL DEFIBRILLATION
Generic Name74DRX
Product CodeDRL
Date Received2009-03-06
Model NumberNA
Catalog Number805249-06
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPHYSIO-CONTROL, INC.
Manufacturer AddressREDMOND WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-03-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.