DISCOVERY DISCOVERY A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-03-10 for DISCOVERY DISCOVERY A manufactured by Hologic, Inc..

Event Text Entries

[17868883] Customer called to report that the unit is making clicking noise and that the x-ray light stays on after the scan has ended. There was no injury to pt or to the operator reported.
Patient Sequence No: 1, Text Type: D, B5


[17910220] After the report, the customer was instructed to remove the unit from use and service engineer was dispatched to perform evaluation and repair of the unit. The engineer confirmed the failure, determined that the failure was caused by a defective printed circuit board. The field engineer replaced the board and verified proper operation of the device. Once the unit was repaired it was returned to customer was use.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1221300-2009-00001
MDR Report Key1377750
Report Source05
Date Received2009-03-10
Date of Report2009-03-06
Date of Event2009-02-11
Date Mfgr Received2008-02-11
Date Added to Maude2009-09-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLIDA REED, DIRECTOR
Manufacturer Street35 CROSBY DR.
Manufacturer CityBEDFORD MA 01730
Manufacturer CountryUS
Manufacturer Postal01730
Manufacturer Phone7819997589
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDISCOVERY
Generic NameBONE DENSITY
Product CodeKGI
Date Received2009-03-10
Model NumberDISCOVERY A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC, INC.
Manufacturer AddressBEDFORD MA US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-03-10

© 2026 FDA.report