ELITE SYSTEM M3 AUTOCLAVABLE FOROBLIQUE TELESCOPE M3-30A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2009-03-24 for ELITE SYSTEM M3 AUTOCLAVABLE FOROBLIQUE TELESCOPE M3-30A manufactured by Gyrus Acmi, Inc. - Norwalk.

Event Text Entries

[1195338] Two devices were used in this procedure, please see medwatch report # 1519132-2009-00006 for the second device. During a prostate surgery, the end of the scope and the tip of the sheath blew apart. Chips of beak were removed from the bladder and the prostate was burned. A new scope and a new sheath were used to finish the procedure.
Patient Sequence No: 1, Text Type: D, B5


[8332890] The customer's observation was not confirmed. The investigation revealed that the distal end of the scope is severly burned. The burn damaged the optics and broke the seal, which allowed moisture into the scope. The burn damage was user error. Another device was also used in this procedure, medwatch report# 1519132-2009-00006 for an emrs-28b sheath. The distal tip of the sheath was broken into several pieces and had burn damage as well. The burn damage was user error.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1519132-2009-00005
MDR Report Key1387717
Report Source07
Date Received2009-03-24
Date of Report2009-03-04
Date of Event2009-03-04
Date Facility Aware2009-03-04
Date Mfgr Received2009-03-06
Device Manufacturer Date2008-10-15
Date Added to Maude2010-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactTERRENCE SULLIVAN
Manufacturer Street136 TURNPIKE RD
Manufacturer CitySOUTHBOROUGH MA 01772
Manufacturer CountryUS
Manufacturer Postal01772
Manufacturer Phone5088042739
Manufacturer G1GYRUS ACMI, INC. - NORWALK
Manufacturer Street93 NORTH PLEASANT STREET
Manufacturer CityNORWALK OH 44857040
Manufacturer CountryUS
Manufacturer Postal Code44857 0409
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELITE SYSTEM M3 AUTOCLAVABLE FOROBLIQUE TELESCOPE
Generic NameM3 AUTOCLAVABLE FOROBLIQUE TELESCOPE
Product CodeFBP
Date Received2009-03-24
Returned To Mfg2009-03-09
Model NumberM3-30A
Catalog NumberM3-30A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age5 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC. - NORWALK
Manufacturer AddressNORWALK OH US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-03-24

© 2026 FDA.report