BACT/ALERT SN CULTURE BOTTLE 259790

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,07 report with the FDA on 2009-05-22 for BACT/ALERT SN CULTURE BOTTLE 259790 manufactured by Biomerieux, Inc..

Event Text Entries

[15697570] There was small hole in the side of a plastic bactalert sn bottle. Since the hole was covered by the product label, it was not seen on visual examination; and the bottle was inoculated and incubated. During incubation, blood and medium oozed through the hole saturating the label and dripping into the cell of the incubator.
Patient Sequence No: 1, Text Type: D, B5


[15890835] An investigation of this incident has been initiated. There were no adverse patient or healthcare worker events associated with this incident. However, there is the potential for adverse events were this to recur since the exposure to some samples can have significant health effects. A domino laser etches the lot id and date code on the label. This laser usually has an exposure time of less than 200 milliseconds. However it is possible for the labeler to stop, positioning a bottle in front of the laser. When this occurs, the laser fires a "standby pulse" into the same location for the duration of the pause. This pulse is capable of burning through the bottle if stationary for more than a few seconds. By turning the standby pulse off, the probabiliity of the laser burning a hole in the bottle should be zero. Biomerieux is also investigating a mechanical shield that would drop into place when the labeler turret is not spinning thus preventing an integrity-compromising hole.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2009-00002
MDR Report Key1391909
Report Source01,05,06,07
Date Received2009-05-22
Date of Report2009-05-21
Date of Event2009-04-23
Date Mfgr Received2009-04-23
Device Manufacturer Date2008-12-31
Date Added to Maude2009-07-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. ELIZABETH FRITZ
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM, NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer Phone9196202682
Manufacturer G1BIOMERIEUX, INC.
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM, NC 27712
Manufacturer CountryUS
Manufacturer Postal Code27712
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBACT/ALERT SN CULTURE BOTTLE
Generic NameMICROBIAL GROWTH MONITOR
Product CodeJTA
Date Received2009-05-22
Catalog Number259790
Lot Number1022154
Device Expiration Date2009-12-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC.
Manufacturer Address100 RODOLPHE STREET DURHAM, NC 27712 US 27712


Patients

Patient NumberTreatmentOutcomeDate
10 2009-05-22

© 2026 FDA.report