VITROS 5,1 FS CHEMISTRY SYSTEM 6801375

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-03-12 for VITROS 5,1 FS CHEMISTRY SYSTEM 6801375 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[1197388] The customer obtained negatively biased phyt results from a single pt sample using vitros phyt slides on a vitros 5,1 fs chemistry system. Biased results of the direction and magnitude observed may lead to inappropriate physician action. The biased results were reported, however, there were no allegations of harm. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[8331691] The investigation determined that phyt quality control results were acceptable at the time of the event. The investigation found no evidence to suggest that either the vitros 5,1 or the vitros phyt slides had malfunctioned. The root cause of the biased phyt result could not be determined, however, pre-analytical error resulting in a short filled sample tube and poor sample quality cannot be ruled out.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319681-2009-00061
MDR Report Key1396813
Report Source05
Date Received2009-03-12
Date of Report2009-02-12
Date of Event2009-02-11
Date Mfgr Received2009-02-12
Device Manufacturer Date2005-09-01
Date Added to Maude2010-02-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOE FALVO
Manufacturer Street100 INDIGO CREEK DR.
Manufacturer CityROCHESTER NY 146265101
Manufacturer CountryUS
Manufacturer Postal146265101
Manufacturer Phone5854535735
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS 5,1 FS CHEMISTRY SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeDKH
Date Received2009-03-12
Model NumberNA
Catalog Number6801375
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DR. ROCHESTER NY 14626510 US 14626 5101


Patients

Patient NumberTreatmentOutcomeDate
10 2009-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.