BIRD GRAPHICS MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2009-06-19 for BIRD GRAPHICS MONITOR manufactured by Cardinal Health.

Event Text Entries

[1101462] Respiratory therapist placed infant on a bvip ventilator. The ventilator and the ventilator alarms were functioning. The graphic monitor was not functioning. Respiratory therapist attempted to check the electric source for the graphic monitor when there was a spark and then smoke. The infant was immediately removed from the ventilator and ventilated with a bag valve mask. A new ventilator was obtained, all systems checked and functional and the infant was placed on ventilator. No adverse outcome to infant or other infants/staff present in the nicu. Biomedical stated where the power cord goes into the fuse block the fuse block cracked in half causing the two wires to short together.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5011626
MDR Report Key1407138
Date Received2009-06-19
Date of Report2009-06-19
Date of Event2009-06-15
Date Added to Maude2009-06-30
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameBIRD GRAPHICS MONITOR
Generic NameMONITOR
Product CodeBZQ
Date Received2009-06-19
Returned To Mfg2009-06-17
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerCARDINAL HEALTH
Manufacturer Address1100 BIRD CENTER DR PALM SPRINGS CA 92262 US 92262


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2009-06-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.