N/A 3797

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1992-09-22 for N/A 3797 manufactured by Davol.

Event Text Entries

[15711679] During laporoscopic cholecystectomy the davol suction irrigator, attached to chemtron flo regular and abbott irrigation fluid bags and had been working properly, ceased to function properly. Saline was not coming out of the port. Surgical team noted that air/nitrogen was coming the port. The alarm for the insufflator went off at this time. The crna noted that abdominal pressure reading at 60 instead of 15. Abdomen immediately desufflated with no adverse clinical effectsdevice labeled for single use. Patient medical status prior to event: unknown. There was not multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. No imminent hazard to public health claimed. Device used as labeled/intended. Device was evaluated after the event. Method of evaluation: performance tests performed. Results of evaluation: component failure. Conclusion: device failure directly caused event. Certainty of device as cause of or contributor to event: yes. Corrective actions: device returned to manufacturer/dealer/distributor. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1410
MDR Report Key1410
Date Received1992-09-22
Date of Report1992-09-10
Date of Event1992-08-06
Date Facility Aware1992-08-06
Report Date1992-09-10
Date Reported to FDA1992-09-10
Date Reported to Mfgr1992-08-06
Date Added to Maude1992-09-29
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameN/A
Generic NameENDOSCOPIC IRRIGATION SET
Product CodeGBQ
Date Received1992-09-22
Model Number3797
Catalog NumberN/A
Lot Number21BC4L22
ID NumberN/A
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Age01-JUL-92
Implant FlagN
Device Sequence No1
Device Event Key1353
ManufacturerDAVOL


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1992-09-22