BACT/ALERT FN CULTURE BOTTLE 259790

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2009-03-30 for BACT/ALERT FN CULTURE BOTTLE 259790 manufactured by Biomerieux, Inc..

Event Text Entries

[1266216] An incubator module rack was contaminated by spilled medium from a positive culture bottle which contained serratia spp. It was reported that the aluminum crimp seal was missing from the culture bottle and the rubber stopper was dislodged. There was no adverse event associated with this incident.
Patient Sequence No: 1, Text Type: D, B5


[8527365] An investigation of this incident has been initiated. The customer alleges that the aluminum crimp seal was missing prior to use. The complaint states that the user removed the flip cap and inoculated the bottle through the rubber stopper without noticing that the aluminum seal was missing. This is an unlikely scenario since the flip top is attached to the crimp seal which covers the rubber stopper. The picture provided by the user shows the rubber stopper with "dents" or gauges in the rubber stopper. The instructions for use advise to visually inspect the bottles prior to use and not to use if there is evidence of damage, leakage or deterioration.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2009-00001
MDR Report Key1418559
Report Source06
Date Received2009-03-30
Date of Report2009-03-16
Date of Event2009-03-02
Date Mfgr Received2009-03-02
Device Manufacturer Date2008-08-01
Date Added to Maude2010-05-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELIZABETH FRITZ
Manufacturer Street100 RODOLPHE ST
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer Phone9196202682
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBACT/ALERT FN CULTURE BOTTLE
Generic NameMICROBIAL GROWTH MONITOR
Product CodeJTA
Date Received2009-03-30
Catalog Number259790
Lot Number1021290
Device Expiration Date2009-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC.
Manufacturer Address100 RODOLPHE ST DURHAM NC US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.