[1266779] It was reported that the wall stand bucky holder assembly fell from approximately 2 feet above the ground and contacted the floor. Prior to the incident, no patient was in the room and the technologist near the wall stand was setting up the assembly. When the technologist turned around, the assembly fell. There was no injury sustained as there was no contact with the technologist. Due to the weight of the assembly, the concern is for a serious injury if the assembly were to fall and contact a patient or technologist.
Patient Sequence No: 1, Text Type: D, B5


[8271294] The assembly is mounted by small metric screws on the top and bottom of the bucky tray assembly. During inspection, the ge field engineer (fe) found that the mounting screws at the bottom of the assembly had come loose and broke. The fe also inspected the assembly and found that the bottom trim cover was damaged as a result of the fall. The fe replaced both the bottom trim cover and the mounting screws.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2126677-2009-00031
MDR Report Key1426096
Report Source05
Date Received2009-04-30
Date of Report2009-04-01
Date of Event2009-03-30
Date Mfgr Received2009-04-01
Device Manufacturer Date1997-04-01
Date Added to Maude2009-09-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARTHA KAMROW
Manufacturer Street3000 N GRANDVIEW BLVD. W450
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal53188
Manufacturer Phone2623127196
Manufacturer G1GE MEDICAL SYSTEMS, LLC
Manufacturer Street3000 N GRANDVIEW BLVD.
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal Code53188
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOMPAX 40E
Product CodeIZZ
Date Received2009-04-30
Model Number2146726
Catalog NumberNA
Lot NumberNA
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGE MEDICAL SYSTEMS, LLC
Manufacturer Address3000 N GRANDVIEW BLVD. WAUKESHA WI 53188 US 53188


Patients

Patient NumberTreatmentOutcomeDate
10 2009-04-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.