FRESENIUS 2008 BSS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1994-03-18 for FRESENIUS 2008 BSS manufactured by Fresenius Usa, Inc..

Event Text Entries

[10497] Pt denied complaints pre-dialysis. Temp. 97. 6, bp 120/70 weight 138 lbs,, gained 2 lbs. Dialysiis was initiated in the usual manner at 300 blood flow rate. 470 ultraaafiltration gaaol 1400cc. 2hrs 5 mins into dialysis pt's bp was 90/60. Pt was lethargicand 200cc normal saline was given. Bp remained low - 80/60-90/60 and dialysis wa terminated at 2hrs 35 mins because blood pressure did not improve in spite of an additional 550cc saline. Pt also reported cramping. Post dialysis temp. 98. 0, bp 120/70, post weight 135 - loss 3 lbs. The pt was asymptomatic post treatment and was sent home. This was the second incident on this machine that involved this pt. It was the third incident on the machine since 11/15/93. The chieeef technician removed the machine from use. The u. F. Rate was recalibrated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2243621-1994-00015
MDR Report Key14307
Date Received1994-03-18
Date of Report1994-01-26
Date of Event1993-11-22
Date Facility Aware1994-01-25
Report Date1994-02-01
Date Reported to FDA1994-02-18
Date Reported to Mfgr1994-02-01
Date Added to Maude1994-06-27
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameFRESENIUS 2008 BSS
Generic NameBEDSIDEE STATIOIN DIALYSIS MACHINE
Product CodeFLC
Date Received1994-03-18
Model Number2008 BSS
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age*
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key14303
ManufacturerFRESENIUS USA, INC.
Manufacturer Address2637 SHADELANDS DRIVE WALNUT CREEK CA 94598 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1994-03-18

© 2026 FDA.report