J.L. SHEPHERD BLOOD IRRADIATOR 143-68A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2009-08-06 for J.L. SHEPHERD BLOOD IRRADIATOR 143-68A manufactured by J.l. Shepherd & Associates.

Event Text Entries

[1190167] The j. L. Shepherd blood irradiator is utilized to irradiate blood and blood products. In order to provide uniform exposure to the irradiation source and thus ensure a predictable dose, the canister containing the blood unit-s rotates during the irradiation cycle. Rotation is confirmed by a light on the device instrument panel. This single visual reference is the only indication that the canister is rotating. There is no alarm that designates failure mode and the device has no computer/software interface to record and monitor device output or flag a mechanical failure of any nature. This compels users to manually record all activity and assures that intermittent failures i. E. , initial rotation followed by one or more failures during a cycle - may not be noticed by a user who has not been able to observe the light continuously during operation. Failure to properly irradiate a blood component may have serious consequences for patients who require this type of blood product. The manufacturer should upgrade all devices to provide at least an audible alarm or computer generated message/alert to designate serious mechanical failure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5012322
MDR Report Key1436254
Date Received2009-08-06
Date of Report2009-08-06
Date of Event2009-08-06
Date Added to Maude2009-08-14
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameJ.L. SHEPHERD BLOOD IRRADIATOR
Generic NameBLOOD IRRADIATOR
Product CodeMOT
Date Received2009-08-06
Model Number143-68A
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerJ.L. SHEPHERD & ASSOCIATES
Manufacturer Address1010 ARROYO AVE. SAN FERNANDO CA 91340182 US 91340 1822


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2009-08-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.