[20469300] Underdelivery report number 2 of 2 for this device. The device tested fine after the first reported incident, thus it was suspected to be a user issue. The pump was then placed again in home care use infusing a tpn solution of 1250ml over 12 hrs. A flow detector was not in use. The pump was set up and sounded" like it was infusing and the display indicated volumes delivered. The infusion was to run overnight. Upon awakening in the morning, the pt noted that the display indicated the full volume had infused, however, the iv container was still almost full.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2921482-1998-00001
MDR Report Key143979
Report Source05
Date Received1998-01-10
Date of Report1997-12-10
Date of Event1997-11-08
Date Mfgr Received1997-12-10
Device Manufacturer Date1995-08-01
Date Added to Maude1998-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameLIFECARE 175 CONTROLLER
Generic NamePUMP CONTROLLER
Product CodeLDR
Date Received1998-01-10
Model NumberNA
Catalog Number11062
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key140407
ManufacturerABBOTT LABORATORIES
Manufacturer Address755 JARVIS DR. MORGAN HILL CA 95037 US
Baseline Brand NameLIFECARE 175 CONTROLLER
Baseline Generic NamePUMP CONTROLLER
Baseline Model NoNA
Baseline Catalog No11062
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1998-01-10