ALCON TOTAL PLUS 23 GA VITRECTOMY PAK 8065750828

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2009-08-20 for ALCON TOTAL PLUS 23 GA VITRECTOMY PAK 8065750828 manufactured by Alcon Laboratories, Inc..

Event Text Entries

[18910711] During a pars plana vitrectomy for repair of a rhegmatogenous retinal attachment, it was noted that the vitrectomy probe became fused to the trocar through which it is placed into the eye. Attempted removal of the probe form the trocar resulted in the outer plastic portion of the trocar separating from the inner metal portion through which the probe is placed. The vitrector was activated to clear the port and then the probe/trocar complex was removed from the eye. Inspection of the retina noted a giant retinal tear, presumably due to vitreous incarceration between the probe and trocar. Dr considers this event to be serious and to have potential to harm pts. The tear was repaired to avoid permanent harm with recovery to be assessed at 12 weeks.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5012491
MDR Report Key1445404
Date Received2009-08-20
Date of Report2009-07-31
Date of Event2009-07-29
Date Added to Maude2009-08-28
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameALCON TOTAL PLUS 23 GA VITRECTOMY PAK
Generic NameVITRECTOMY PAK
Product CodeMSR
Date Received2009-08-20
Catalog Number8065750828
Lot NumberBOX NOT SAVED
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerALCON LABORATORIES, INC.
Manufacturer Address6201 SOUTH FREEWAY FORT WORTH TX 76134200 US 76134 2009


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2009-08-20