PSS SELECT STREP A DIPSTICK IST-501-OBC507

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07,08 report with the FDA on 2009-06-01 for PSS SELECT STREP A DIPSTICK IST-501-OBC507 manufactured by Biosite Inc..

Event Text Entries

[21324967] Investigation: product lot: sta7080127 (exp. Date: 2009/08). Summary of results: according to the complaint info, retained strips were tested with qc positive standards and positive control; all strips yielded positive results and met qc specifications. The results above indicate that the complaint was not confirmed. It is possible that the strep a present was below the detectable level of the test, which may cause a negative result. The product number, product description, lot number and batch record were verified. No abnormal results were found in the batch record review. No corrective action required as no product deficiency was established.
Patient Sequence No: 1, Text Type: N, H10


[21426974] In 2009- customer complained of false negative results on pss strep a test. It showed positive only after 1 to 2 hours. Various pt swabs were compared to a culture sent to the lab, and the lab results came back as positive.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2009-00381
MDR Report Key1446094
Report Source07,08
Date Received2009-06-01
Date of Report2009-06-01
Date of Event2009-05-05
Date Mfgr Received2009-05-05
Date Added to Maude2009-10-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMEN BERGELIN, MGR.
Manufacturer Street9975 SUMMERS RIDGE RD.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052256
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNI
Event Type3
Type of Report3

Device Details

Brand NamePSS SELECT STREP A DIPSTICK
Generic NameSTREP A TEST
Product CodeGTY
Date Received2009-06-01
Model NumberIST-501-OBC507
Lot NumberSTA7080127
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOSITE INC.
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-06-01

© 2026 FDA.report