MTS ANTI-IGG, -C3D CARD MTS4014

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2009-05-08 for MTS ANTI-IGG, -C3D CARD MTS4014 manufactured by Micro Typing Systems, Inc,.

Event Text Entries

[1153631] The customer reported that a dat positive sample failed to react in the mts anti-igg, -c3d card lot# 071608005-04. The sample reacted positive (1+) using the mts anti-igg card. A false negative dat may lead to confusion in the diagnosis of anemia or misdiagnosis of a hemolytic transfusion reaction.
Patient Sequence No: 1, Text Type: D, B5


[8310957] Satisfactory results were obtained at ocd using retained cards from mts anti-igg - c3d card lot # 071608005-04 and mts anti-igg card lot # 112408001-30 using known dat positive and negative samples. The gel cards performed as expected at ocd. The customer's complaint was not confirmed. The sample in question was submitted to ocd for add'l testing. A definite root cause was not determined regarding the negative results obtained at the customer site. A complaint review indicated no other complaints have been logged against lot# 071608005-04. Batch record review was within release specifications. Incident is isolated. (b) (4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2009-00097
MDR Report Key1458586
Report Source05,06
Date Received2009-05-08
Date of Report2009-05-07
Date of Event2009-04-14
Date Mfgr Received2009-04-14
Device Manufacturer Date2008-11-01
Date Added to Maude2010-05-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJENNIFER SPARROW
Manufacturer Street1001 US ROUTE 202
Manufacturer CityRARITAN NJ 08869
Manufacturer CountryUS
Manufacturer Postal08869
Manufacturer Phone9087043872
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS ANTI-IGG, -C3D CARD
Generic NameANTI-HUMAN GLOBULIN REAGENT
Product CodeCFQ
Date Received2009-05-08
Model NumberNA
Catalog NumberMTS4014
Lot Number071608005-04
ID NumberNA
Device Expiration Date2009-08-24
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMICRO TYPING SYSTEMS, INC,
Manufacturer Address1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US 33069