MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07 report with the FDA on 2009-09-04 for BECKMAN COULTER OLA2500 LAB AUTOMATION NA manufactured by Olympus Life Science Research Europa.

Event Text Entries

[16035558] Complaint was received from outside the usa. The customer reported an aliquot tube with two different barcode labels.
Patient Sequence No: 1, Text Type: D, B5


[16115218] No information has been received to suggest any ill effects associated with this event. Double labeling occurred on an aliquot tube because the operator did not follow instructions to check the label roll in the printer and remove the first label when an on-screen error message appeared. Both labels were affixed to the same aliquot tube. The printer was investigated by the manufacturer and no obvious error could be detected. If the investigation reveals significant new information, a follow-up will be submitted. Labeling in the (b)(4) user guide states action to be followed in the event a printer error message occurs and is considered adequate. Beckman coulter, inc. Is not aware of any incidence of this type reported in the (b)(6).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006655511-2009-00001
MDR Report Key1466352
Report Source01,06,07
Date Received2009-09-04
Date of Report2009-09-04
Date of Event2009-08-07
Date Mfgr Received2009-08-07
Device Manufacturer Date2005-02-01
Date Added to Maude2011-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID DAVIS
Manufacturer Street4300 N. HARBOR BLVD.
Manufacturer CityFULLERTON CA 92835
Manufacturer CountryUS
Manufacturer Postal92835
Manufacturer Phone9726194710
Manufacturer G1OLYMPUS LIFE SCIENCE RESEARCH EUROPA GMBH
Manufacturer StreetRUPPERT-MAYER-STRASSE 44
Manufacturer CityMUENCHEN 81379
Manufacturer CountryGM
Manufacturer Postal Code81379
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBECKMAN COULTER OLA2500 LAB AUTOMATION
Generic NameLAB AUTOMATION
Product CodeJJG
Date Received2009-09-04
Model NumberOLA2500
Catalog NumberNA
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS LIFE SCIENCE RESEARCH EUROPA
Manufacturer AddressMUNICH GM


Patients

Patient NumberTreatmentOutcomeDate
10 2009-09-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.