SMART MONITOR *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2009-08-26 for SMART MONITOR * manufactured by Respironics, Inc..

Event Text Entries

[1152510] While the mother of the infant was feeding the baby, the apnea monitor respiration light stopped blinking and the heart rate light started to slow. The baby stopped breathing. The baby's family performed cpr and the baby was transported to the hospital for observation. The patient's family noted the heart rate was greater than 200 bpm, but noted the monitor did not alarm. However, the high heart rate indicator light was showing red; indicating that an alarm had occurred. The alarm settings were hhr=200, lhr=80, apnea delay= 20 seconds. The monitor was removed from the home and sent to the manufacturer for testing. Initial response has been no problems identified with the monitor; facility is currently awaiting written report. They could see two events, and according to the read out, the alarm sounded both times.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1472619
MDR Report Key1472619
Date Received2009-08-26
Date of Report2009-08-26
Date of Event2009-08-06
Report Date2009-08-26
Date Reported to FDA2009-08-26
Date Added to Maude2009-09-15
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSMART MONITOR
Generic NameAPNEA MONITOR
Product CodeNPF
Date Received2009-08-26
Model Number*
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerRESPIRONICS, INC.
Manufacturer Address1001 MURRY RIDGE LANE MURRYSVILLE PA 15668 US 15668


Patients

Patient NumberTreatmentOutcomeDate
10 2009-08-26