THERMOCHEM HT-1000 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2009-09-01 for THERMOCHEM HT-1000 * manufactured by Thermasolutions.

Event Text Entries

[1132871] Thermocare machine was being used with a hipec (heated intraoperative intraperitoneal chemotherapy) case. The perfusionist said the screen froze. Temperature regulation via the machine was disabled and manual temperature regulation measures were used to proceed with the case. There was no effect/injury to the patient. Our biomed department did not assess the device as it was returned to the manufacturer's representative immediately after the case. (this is not our procedure and that has been addressed at our level). Biomed reports that this is the third time in four cases that we have had problems with an ht-1000 from thermasolutions. All the occurrences indicate screen issues. However, there are no other similar devices available per biomed department.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1478197
MDR Report Key1478197
Date Received2009-09-01
Date of Report2009-09-01
Date of Event2009-08-17
Report Date2009-09-01
Date Reported to FDA2009-09-01
Date Added to Maude2009-09-22
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameTHERMOCHEM
Generic NameINTRAPERITONEAL HYPERTHERMIA SYSTEM
Product CodeMLW
Date Received2009-09-01
Returned To Mfg2009-08-17
Model NumberHT-1000
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerTHERMASOLUTIONS
Manufacturer Address1889 BUERKLE ROAD WHITE BEAR LAKE MN 55110 US 55110


Patients

Patient NumberTreatmentOutcomeDate
10 2009-09-01