MEDGRAPHICS 1085-D

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1998-01-14 for MEDGRAPHICS 1085-D manufactured by Medgraphics.

Event Text Entries

[108107] In 2/97, three medgraphics 1085 advantage plus systems were purchased by medical center to replace the older 1070 models. In 6/97, it was noted that the 1085 was producing comparative diffusion readings 5-10% lower than the 1070. In 11/97, g. C. Columns were replaced in the three 1085 units. The 1085 units continued to produce results that were lower than the 1070 beyond the acceptable variation (6%).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number148756
MDR Report Key148756
Date Received1998-01-14
Date of Report1997-12-18
Date Facility Aware1997-11-01
Report Date1997-12-18
Date Reported to Mfgr1997-12-18
Date Added to Maude1998-02-17
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameMEDGRAPHICS
Generic NameBODY PLETHYSMOGRAPH WITH DIFFUSING CAPACITY
Product CodeCCM
Date Received1998-01-14
Model Number1085-D
Catalog NumberNI
Lot NumberNI
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age11 MO
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key144960
ManufacturerMEDGRAPHICS
Manufacturer Address350 OAK GROVE PKWY ST. PAUL MN 55127 US

Device Sequence Number: 2

Brand NameMEDGRAPHICS
Generic NameBODY PLETHYSMOGRAPH WITH DIFFUSING CAPACITY
Product CodeCCM
Date Received1998-01-14
Model Number1085-D
Catalog NumberNI
Lot NumberNI
ID NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age11 MO
Implant FlagN
Date RemovedA
Device Sequence No2
Device Event Key144961
ManufacturerMEDGRAPHICS
Manufacturer Address350 OAK GROVE PKWY ST. PAUL MN 55127 US

Device Sequence Number: 3

Brand NameMEDGRAPHICS
Generic NameBODY PLETHYSMOGRAPH WITH DIFFUSING CAPACITY
Product CodeCCM
Date Received1998-01-14
Model Number1085-D
Catalog NumberNI
Lot NumberNI
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age11 MO
Implant FlagN
Date RemovedA
Device Sequence No3
Device Event Key144964
ManufacturerMEDGRAPHICS
Manufacturer Address350 OAK GROVE PKWY ST. PAUL MN 55127 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1998-01-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.