MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1998-01-14 for MEDGRAPHICS 1085-D manufactured by Medgraphics.
[108107]
In 2/97, three medgraphics 1085 advantage plus systems were purchased by medical center to replace the older 1070 models. In 6/97, it was noted that the 1085 was producing comparative diffusion readings 5-10% lower than the 1070. In 11/97, g. C. Columns were replaced in the three 1085 units. The 1085 units continued to produce results that were lower than the 1070 beyond the acceptable variation (6%).
Patient Sequence No: 1, Text Type: D, B5
Report Number | 148756 |
MDR Report Key | 148756 |
Date Received | 1998-01-14 |
Date of Report | 1997-12-18 |
Date Facility Aware | 1997-11-01 |
Report Date | 1997-12-18 |
Date Reported to Mfgr | 1997-12-18 |
Date Added to Maude | 1998-02-17 |
Event Key | 0 |
Report Source Code | User Facility report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MEDGRAPHICS |
Generic Name | BODY PLETHYSMOGRAPH WITH DIFFUSING CAPACITY |
Product Code | CCM |
Date Received | 1998-01-14 |
Model Number | 1085-D |
Catalog Number | NI |
Lot Number | NI |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | 11 MO |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 1 |
Device Event Key | 144960 |
Manufacturer | MEDGRAPHICS |
Manufacturer Address | 350 OAK GROVE PKWY ST. PAUL MN 55127 US |
Brand Name | MEDGRAPHICS |
Generic Name | BODY PLETHYSMOGRAPH WITH DIFFUSING CAPACITY |
Product Code | CCM |
Date Received | 1998-01-14 |
Model Number | 1085-D |
Catalog Number | NI |
Lot Number | NI |
ID Number | NI |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | 11 MO |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 2 |
Device Event Key | 144961 |
Manufacturer | MEDGRAPHICS |
Manufacturer Address | 350 OAK GROVE PKWY ST. PAUL MN 55127 US |
Brand Name | MEDGRAPHICS |
Generic Name | BODY PLETHYSMOGRAPH WITH DIFFUSING CAPACITY |
Product Code | CCM |
Date Received | 1998-01-14 |
Model Number | 1085-D |
Catalog Number | NI |
Lot Number | NI |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | 11 MO |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 3 |
Device Event Key | 144964 |
Manufacturer | MEDGRAPHICS |
Manufacturer Address | 350 OAK GROVE PKWY ST. PAUL MN 55127 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 1998-01-14 |