PROBECHEK PATHVYSION HER-2/NEU NORMAL CONTROL 02J05-030

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-09-01 for PROBECHEK PATHVYSION HER-2/NEU NORMAL CONTROL 02J05-030 manufactured by Abbott Molecular, Inc..

Event Text Entries

[1173882] A customer reported that probechek pathvysion control slides that were shipped were broken upon arrival. The plastic case housing the probechek pathvysion control slides had popped open. The customer stated two problems with the box's contents: heavy reagents were on top of the probechek slides, no cushion wrap was used around the slides during shipment. The slides were broken. Customer stated that a person could have cut themselves due to the glass shards. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[8284172] Abbott molecular at this time believes that this incident is a packaging error - i. E. , the customer reported that no packaging material was used to wrap the box of the probechek slides for the pathvysion her-2/neu assay. The review of the operating procedures for the packaging of the slides was completed and determined that only cold storage boxes of slides are cushioned wrapped, but not for ambient shipment.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005248192-2009-00003
MDR Report Key1497200
Report Source05
Date Received2009-09-01
Date of Report2009-08-31
Date of Event2009-08-04
Date Mfgr Received2009-08-04
Date Added to Maude2010-08-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactPAULA MARTIN
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2243617333
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROBECHEK PATHVYSION HER-2/NEU NORMAL CONTROL
Generic NameMICROSCOPIC SLIDES USED IN FISH TESTING
Product CodeKIR
Date Received2009-09-01
Catalog Number02J05-030
Lot NumberTBD
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer Address1300 EAST TOUHY AVENUE DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2009-09-01

© 2026 FDA.report