BULLSEYE FEMORAL AIMER 6MM C8660

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07 report with the FDA on 2009-09-03 for BULLSEYE FEMORAL AIMER 6MM C8660 manufactured by Conmed Linvatec.

Event Text Entries

[1241345] The user reported that during use of this guide for an acl procedure, the tip broke off in the surgical site. The tip was retrieved and the surgeon completed the procedure using another femoral guide. There was no pt injury or significant surgical delay related to this event.
Patient Sequence No: 1, Text Type: D, B5


[8290643] Conmed linvatec received this device for eval. An investigation of this failure remains in process. A f/u report will be sent upon completion of the investigation. The info for use (ifu) provides the following precautions: inspect instrument prior to use to ensure it is in good physical condition and functions properly. There should be no loose, broken or misaligned parts. Exercise care in the use of the device to minimize side or bending loads. Do not use excessive force on instruments to avoid damage or breakage during use. Inspect instrument after use to ensure it has not been damaged.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017294-2009-00129
MDR Report Key1497516
Report Source01,06,07
Date Received2009-09-03
Date of Report2009-06-26
Date of Event2009-06-26
Date Mfgr Received2009-06-26
Device Manufacturer Date2008-12-01
Date Added to Maude2010-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBEVERLY SCHANER
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal33773
Manufacturer Phone7273995169
Manufacturer G1CONMED LINVATEC
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal Code33773
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBULLSEYE FEMORAL AIMER 6MM
Generic NameGUIDE, SURGICAL, INSTRUMENT
Product CodeLXI
Date Received2009-09-03
Returned To Mfg2009-09-02
Model NumberNA
Catalog NumberC8660
Lot Number10516A
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCONMED LINVATEC
Manufacturer AddressLARGO FL US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-09-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.