PSS CONSULT HCG URINE CASSETTE 4584655014

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2009-09-18 for PSS CONSULT HCG URINE CASSETTE 4584655014 manufactured by Biosite Incorporated.

Event Text Entries

[1222676] Caller reported false negative urine hcg result. No quantitative hcg result provided. The pregnancy was confirmed by ultrasound - 9+ weeks.
Patient Sequence No: 1, Text Type: D, B5


[8283835] Investigation: lot number(s): hcg9030206. Expiration date(s): 03/2011. Control lot: 20miu/ml hcg urine - lot: hcg090304-01; 100miu/ml hcg urine - lot: hcg090521-01; 255. 3iu/ml hcg urine - lot: hcg090526-03; 500iu/ml hcg - lot: hcg090603-01. Summary of results: the retention devices meet qc specification. Correct positive results were observed when tested with 20miu/ml and 100miu/ml hcg urine control at 3 minutes read time. The high hcg samples (255. 3iu/ml hcg and 500iu/ml hcg) yielded clearly positive results at 3 minutes reading time. Conclusion: the retention devices meet qc specification. No abnormal result was observed; this complaint is not confirmed. Nothing abnormal found in batch review and the product number, product description and lot number was verified. No corrective action is required at this time as product deficiency was not established.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2009-00839
MDR Report Key1504018
Report Source05,07
Date Received2009-09-18
Date of Report2009-09-18
Date of Event2009-08-26
Date Mfgr Received2009-08-26
Date Added to Maude2009-12-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMEN BERGELIN, MANAGER
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052256
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNI
Event Type3
Type of Report3

Device Details

Brand NamePSS CONSULT HCG URINE CASSETTE
Generic NameHCG PREGNANCY TEST
Product CodeJHJ
Date Received2009-09-18
Model Number4584655014
Lot NumberHCG9030206
Device Expiration Date2011-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOSITE INCORPORATED
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-09-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.