EPICARE COLUMN 680

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-10-08 for EPICARE COLUMN 680 manufactured by Hill-rom Architecural Products / Hill-rom Co. Inc.

Event Text Entries

[17472289] Alleged the retractable ceiling column drifts down unexpectedly. On (b)(6), an or employee was hit on the head when the ceiling column drifted down. The employee had a headache, went to er, is reported to have a concussion and was sent home for the day. The employee is working but still claims to have an ongoing headache. The facility indicated the drift has been happening for sometime.
Patient Sequence No: 1, Text Type: D, B5


[17637815] The technician inspected the column and could not duplicate the alleged malfunction. The technician did find small air leaks between the control hoses and press lock fittings. The hoses were re-fitted to the press lock fittings to stop the leaks. The account declined to provide any additional information regarding the event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1836145-2009-00003
MDR Report Key1506623
Report Source05
Date Received2009-10-08
Date of Report2009-09-09
Date of Event2009-09-08
Date Mfgr Received2009-09-09
Date Added to Maude2010-07-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARK VORNHEDER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129347163
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEPICARE COLUMN
Generic NameCEILING COLUMN
Product CodeIYB
Date Received2009-10-08
Model Number680
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM ARCHITECURAL PRODUCTS / HILL-ROM CO. INC
Manufacturer AddressWEST PEARL & WALNUT STS BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2009-10-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.