PRIMA XTEND ARM 129244

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2009-09-25 for PRIMA XTEND ARM 129244 manufactured by Hill-rom Architectural Products - Hill-rom Co. Inc.

Event Text Entries

[1244030] The facilities engineer alleged the gcx bracket came loose from the prima xtend unit and fell on top of a crib. A pt was not in the crib. The nurse was in the process of getting the crib ready for a baby, and as she moved the monitor, the monitor arm fell on top of the crib. The facilities engineer was able to tighten the two screws on the bracket which secures the monitor arm to the prima xtend. He also checked several of the other units in the area and found loose screws.
Patient Sequence No: 1, Text Type: D, B5


[8397750] The tech investigated and found monitor arm bracket screws came loose. He did not see loctite on the screws and reinstalled the screws using loctite. All units at this account were checked and loctite was added to each of the screws. The tech indicated that the monitors are moved around a lot at this facility and recommended that the facility tighten the screws as part of an annual preventative maintenance program.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1836145-2009-00002
MDR Report Key1507336
Report Source00
Date Received2009-09-25
Date of Report2009-08-27
Date of Event2009-08-26
Date Mfgr Received2009-08-27
Date Added to Maude2010-07-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARK VORNHEDER
Manufacturer Street1069 STATE RT 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129347163
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRIMA XTEND ARM
Generic NameSUSPENSION ARM
Product CodeIYB
Date Received2009-09-25
Model Number129244
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM ARCHITECTURAL PRODUCTS - HILL-ROM CO. INC
Manufacturer AddressWEST PEARL AND WALNUT STS BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2009-09-25

© 2024 FDA.report