ON-BOARD IMAGER (OBI) H08

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2009-09-23 for ON-BOARD IMAGER (OBI) H08 manufactured by Varian Medical Systems.

Event Text Entries

[1178840] It was reported that during a scheduled oncology treatment, while moving the kvd arms into position from the obi console, the kvd panel cover fell off the assembly and struck a pt in the head. The pt was observed by hosp staff and believed not to have encountered any serious injury because of the incident. No further medical attention was required other than the separate observation.
Patient Sequence No: 1, Text Type: D, B5


[8418895] The pt's additional exam did not show evidence of any injury from the impact encountered. In multiple reports of this malfunction, no serious injury has resulted, and no medical intervention beyond exam and/or diagnostic testing has resulted. After the reported occurrence, varian service instructed the hosp's biomedical engineering dept to secure the cover to the imager with screws. This is a varian approved modification which replaces the usage to the originally designed velcro fasteners. Varian has come to a decision to report incidents involving medical intervention (x-ray and ct scans) or observations. Varian has determined that an mdr is appropriate, as this malfunction, should it recur, could potentially cause a serious injury.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916710-2009-00039
MDR Report Key1508175
Report Source01,05,06
Date Received2009-09-23
Date of Report2009-09-08
Date of Event2009-09-08
Device Manufacturer Date2006-11-01
Date Added to Maude2010-02-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVE HALL, ACTING MGR
Manufacturer Street911 HANSEN WAY M/S C-255
Manufacturer CityPALO ALTO CA 94304
Manufacturer CountryUS
Manufacturer Postal94304
Manufacturer Phone6504833153
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameON-BOARD IMAGER (OBI)
Generic NameMEDICAL PARTICAL CHARGED RAD. THERAPY
Product CodeLHN
Date Received2009-09-23
Model NumberH08
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVARIAN MEDICAL SYSTEMS
Manufacturer AddressPALO ALTO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-09-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.